LETTER TO THE EDITOR

Preliminary feasibility of pre-treatment F-18-PSMA-1007 PET/CT in dose prediction for Lu-177-PSMA-I&T therapy

Suvi Kokkonena,b symbol, Vappu Reijonena symbol, Eero Hippeläinenb,c symbol, Veera Ahtiainena symbol, Jukka Schildtc symbol, Outi Sipiläc symbol, Sauli Savolainenb symbol and Mikko Tenhunena symbol

aCancer Center, Helsinki University Hospital, Helsinki, Finland; bDepartment of Physics, University of Helsinki, Helsinki, Finland; cClinical Physiology and Nuclear Medicine, Helsinki University Hospital, Helsinki, Finland

KEYWORDS: Castration-resistant prostate cancer; PSMA-PET; [177Lu]Lu-PSMA radioligand therapy; theranostic

 

Citation: ACTA ONCOLOGICA 2026, VOL. 65, 609–613. https://doi.org/10.2340/1651-226X.2026.45700.

Copyright: © 2026 The Author(s). Published by MJS Publishing on behalf of Acta Oncologica. This is an Open Access article distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/).

Received: 10 March 2026; Accepted: 28 June 2026; Published: 22 July 2026

CONTACT: Suvi Kokkonen suvi.e.kokkonen@hus.fi HUS Cancer Center, Paciuksenkatu 3, 00290 Helsinki, Finland

Supplemental data for this article can be accessed online at https://doi.org/10.2340/1651-226X.2026.45700

Competing interests and funding: VR has received institutional funding for clinical phase-I studies by a sponsor (Bayer) and a dosimetry trial (Pharmtrace). EH receives institutional funding for dosimetry services as part of a clinical trial (Pharmtrace).

 

Introduction

In metastatic castration-resistant prostate cancer (mCRPC), the treatment options are limited to palliative approaches, including targeted radionuclide therapy (TRT) using Lutetium-177 prostate-specific membrane antigen ([177Lu]Lu-PSMA) [1, 2]. Eligibility for [177Lu]Lu-PSMA therapy is commonly assessed with a pre-treatment Fluorine-18 [18F]F-PSMA positron emission tomography/computed tomography (PET/CT) to confirm sufficient PSMA accumulation in tumor lesions [3].

TRT is typically administered with fixed activity, not accounting for interpatient variability in biodistribution and biokinetics, leading to potential under- or overtreatment in some patients [4, 5]. The kidneys are considered critical organs in [177Lu]Lu-PSMA therapies, in addition to red bone marrow, salivary, and lacrimal glands. Patient-specific dosimetry is recommended [6].

There is significant variation in the reported dose estimates, which may be not only related to patient characteristics but also to the influence of heterogenous methodological factors [7]. Post-treatment dosimetry in [177Lu]Lu-PSMA therapy is typically performed using single-photon emission computed tomography/computed tomography (SPECT/CT), but the imaging time points, acquisition and reconstruction protocols, and segmentation methods vary between centers. There are initiatives for harmonization [810]. Several studies have addressed the predictive power of pre-therapy PSMA PET imaging with promising, yet varying results [1113]. No established method exists to use pre-therapy PET imaging for individualized dose planning in PSMA TRT.

Recent studies report correlations between the [177Lu]Lu-PSMA absorbed dose to tumor, the pre-treatment PSMA PET/CT uptakes, and treatment response, suggesting the treatment response could be refined with a dosimetric approach [14, 15]. Preliminarily, we hypothesize that pre-treatment PSMA PET/CT may provide quantitative information to predict absorbed doses. Exploration of the technical factors is a mandatory step preceding this, and a pilot approach was taken.

Patients/material and methods

Patient data

This retrospective feasibility study included five mCRPC patients who underwent both [18F]F-PSMA ([18F]F-PSMA-1007) pre-treatment PET/CT imaging and [177Lu]Lu-PSMA ([177Lu]Lu-PSMA-I&T) therapy followed by two time-point post-treatment SPECT/CT imaging at the Hospital District of Helsinki (HUS).

[177Lu]Lu-PSMA therapy was carried out with an administered activity of 7.4 GBq (± 5%) per cycle. Dosimetry results from the first therapy cycle are studied.

In total, 10 kidney and six bone lesion volumes of interest (VOI) were analyzed. The patient-specific administered activities and CT-based volumes of kidneys and bone lesions are summarized in Table 1. The methodological workflow is presented in Supplementary Appendix A.

Table 1. Administered activities of [18F]F-PSMA for pre-treatment PET/CT and [177Lu]Lu-PSMA treatment, including the CT-based segmented volumes for kidneys, denoted as KL for left and KR for right, and bone lesions, denoted as BL.
[18F]F-PSMA PET/CT
VCT [mL]
[177Lu]Lu-PSMA SPECT/CT
VCT [mL]
Patient # A [MBq] KL KR BL A [GBq] KL KR BL
1 219 147 117 3.8 7.5 163 160 3.7
2 301 249 236 11.2 7.6 258 240 10.4
3 (1) 219 141 131 16.4 7.3 150 164 6.3
3 (2) 9.6 6.6
4 213 121 145 18.0 7.1 132 145 20.8
5 244 116 147 0.5 7.4 119 168 0.9
Median (range) 219 (213–301) 143 (116–249) 10.4 (0.5–18.0) 7.4 (7.1–7.6) 165.5 (119–258) 6.5 (0.9–20.8)
PSMA: Prostate-specific membrane antigen; PET/CT: Positron emission tomography/computed tomography; VCT: CT-based delineated volume; SPECT: Single-photon emission computed tomography; KL: Left kidney; KR: Right kidney; BL: Bone lesion.

Imaging and reconstruction protocols

PET/CT imaging was performed approximately 90 min post [18F]F-PSMA injection. Two SPECT/CT scans were taken approximately 24 h and a week post-treatment [6].

For PET/CT, either GE Healthcare Discovery MI (GE Healthcare, Chicago, IL, USA) or Siemens Biograph mCT 64 R4 (Siemens Healthcare, Erlangen, Germany) was used with EARL2-compliant reconstructions [16]. For SPECT/CT, Siemens SYMBIA pro.Specta Q3 (Siemens Healthcare, Erlangen, Germany) was used with Ordered Subsets Conjugate Gradient Minimizer (OSCGM) reconstruction [17]. Imaging and reconstruction parameters are listed in Appendix B.

Segmentation

The kidneys and bone lesions were segmented separately on the PET/CT and SPECT/CT images, using CT- and gradient-based approaches. CT-based kidney segmentation was performed using MIM Software’s ProtegeAI+ workflow, while lesions were manually segmented on CT [18]. Gradient-based segmentation for both kidneys and lesions was performed using the PET/SPECT intensity gradient-based PET Edge tool in MIM Software [19]. Only lesions clearly visible in PET, SPECT, and CT images were included.

Recovery coefficients

Recovery coefficient (RC) correction was applied to activity uptakes on CT-based bone lesion volumes to improve post-therapy SPECT scan’s quantitative accuracy. RCs for SPECT’s OSCGM reconstruction were determined using the standard International Electrotechnical Commission (IEC) Body Phantom from the National Electrical Manufacturers Association (NEMA) [20]. The phantom spherical inserts were filled with 177Lu water solution (6.6 ± 0.2 MBq/g), followed by SPECT/CT scan using Siemens Pro.Specta Q3 [21]. Detailed imaging parameters are found in Supplementary Appendix B.

Dosimetry and dose prediction

Post-therapy absorbed doses were determined based on two-time-point SPECT/CT imaging and implemented in MIM Software [5].

Pre-treatment dose predictions were derived from PET/CT using an experimental dosimetry module (Madsen Dosimetry Estimator, MIM Software) [22], which extrapolates absorbed dose from pre-therapy activity distributions under specified effective half-life assumptions, detailed below.

Three different effective half-lives for kidneys and bone lesions were tested: patient-specific values observed from post-therapy SPECT/CT dosimetry (CT-based volumes and OSCGM reconstruction), and two literature-based values by Schuchardt et al. (33 h for kidneys and 43 h for bone lesions) [23], and Karimzadeh et al. (39 h for kidneys and 56 h for bone lesions) [24]. The patient-specific post-therapy dosimetry results and effective half-lives were considered as the reference, while literature half-life-based predictions were evaluated against these to assess the feasibility of dose prediction without the prior knowledge of individual kinetics.

Results

Recovery coefficients

SPECT recovery coefficients were modeled using monoexponential fits. For OSCGM reconstruction, the recovery curve was:

RC=1.090-1.607e-0.052D

where D is the sphere diameter (mm). The RC measurements and the model fit are presented in Supplementary Appendix C.

Post-therapy dosimetry using SPECT/CT

The kidneys’ median absorbed dose using OSCGM reconstruction and CT-based segmentation was 3.1 Gy (range 2.0–5.9 Gy). For bone lesions, the median absorbed dose was 33.8 Gy (range 10.4–68.9 Gy). Patient-specific results can be found in Table 2 (under Dref).

Table 2. Absorbed dose predictions (DSchuchardt) derived from pre-treatment PET/CT scan, using CT-based segmentation and literature effective half-lives [23] for kidneys (tSchuchardt = 33 h) and bone lesions (tSchuchardt = 43 h).
Kidneys
PET/CT SPECT/CT Ratio of doses from PET/SPECT
DSchuchardt [Gy] Dref [Gy] tref [h] DSchuchardt/Dref
Patient L R L R L R L R
1 7.5 6.6 3.1 3.1 30 32 2.4 2.1
2 5.0 5.8 2.0 2.4 28 29 2.5 2.4
3 4.5 2.8 3.7 4.3 35 30 1.2 0.7
4 5.5 5.5 2.0 2.1 32 30 2.8 2.6
5 4.6 6.0 4.1 5.9 29 29 1.1 1.0
Median (range) 5.5 (2.8–7.5) 3.1 (2.0–5.9) 30 (29–35) 2.3 (0.7–2.8)
Bone lesions
1 1.5 10.4 67 0.1
2 13.7 21.2 61 0.6
3 (1) 28.1 68.9 159 0.4
3 (2) 14.2 49.2 159 0.3
4 24.6 46.3 61 0.5
5 23.9 16.7 32 1.4
Median (range) 19.1 (1.5–28.1) 33.8 (10.4–68.9) 64 (32–159) 0.5 (0.1–1.4)
PET: Positron emission tomography; SPECT: Single-photon emission tomography.
SPECT/CT-based mean absorbed doses (Dref) and effective half-lives (tref) are reported using CT-based segmentation and OSCGM reconstruction. Activity uptakes on CT-based lesion volumes have been RC corrected for post-therapy dosimetry results. Prediction performance is expressed as the ratio of PET/CT-based predicted doses to SPECT/CT-based observed doses.

Using OSCGM reconstruction and gradient-based segmentation, the kidney median absorbed dose was 2.7 Gy (range 1.8–8.9 Gy), and the median bone lesion absorbed dose was 20.6 Gy (6.5–44.5 Gy).

Pre-therapy dose estimation using PET/CT

Table 2 summarizes the key findings: PET/CT-based absorbed dose predictions for kidneys and bone lesions using literature effective half-lives [23], with SPECT/CT-based observed absorbed doses and corresponding effective half-lives, using current local clinical methodology. Here, the literature half-lives [23] were chosen as the study included more patients (51 patients in [23] vs. 16 in [24]). The median ratio between PET-predicted and SPECT-observed absorbed doses for kidneys was 2.3 (range 0.7–2.8). For lesions, the corresponding median ratio was 0.5 (range 0.1–1.4).

Using the literature effective half-lives [24], the median ratio for kidneys was 2.8 (range 0.8–3.6) and for lesions 0.6 (range 0.2–1.8). Using the reference effective half-lives in PET prediction, the median ratio was 2.3 (range 0.6–3.2) for kidneys and 1.0 (range 0.2–1.5) for lesions.

Dose predictions for combinations of the three effective half-lives and the two segmentation methods are presented in detail in Supplementary Appendix D and E.

Discussion

This pilot study evaluated the feasibility of pre-treatment dosimetry from single time-point [18F]F-PSMA-1007 PET/CT to predict absorbed doses in [177Lu]Lu-PSMA-I&T therapy. The main preliminary finding is that PET-based dose prediction tended to overestimate kidney absorbed dose and underestimate the bone lesion absorbed dose, with substantial interpatient and inter-lesion variability. The segmentation methods produced more consistent dosimetry results for kidneys, while greater variability was observed in lesions. The median absorbed doses were comparable to those reported in previous studies [14, 2328].

Clearly, our pilot approach had limitations. First, the study had a limited patient cohort, and consequently such a limited sample size has no statistical power. The number of analyzed lesions per patient was limited due to the selection criteria, leading to potential high uncertainty. Further studies with a larger sample size are needed.

Furthermore, only the low-dose, noncontrast CT scans were utilized with PET and SPECT imaging. Using diagnostic CT and MRI scans could improve the accuracy of lesion delineation. The volumes of the lesions analyzed were below 21 mL, and the partial volume effect limits accurate activity estimation. Basic RC correction was carried out for post-therapy SPECT/CT scans with diligent registration to counterbalance these effects. PET-based lesion dose predictions could be improved by applying similar corrections. Despite EARL2 reconstructions, the use of two different PET/CT systems may affect the quantitation.

The SPECT-observed effective half-lives for the kidneys aligned more closely with the first set of literature-reported effective half-lives [23] than with the second set [24]. On the other hand, the SPECT-observed mean effective half-lives for bone lesions resembled the second set of literature values [24] more than the first [23]. Especially the SPECT-observed bone lesion effective half-lives exhibited a wide range of values, speaking for the need of using patient- or tumor-specific effective half-lives to achieve better accuracy in dose prediction. The accuracy of post-treatment dosimetry results could be improved using SPECT/CT scans from more time points. The results could also improve, for instance, with population pharmacokinetic modeling [29].

An avenue to explore could be to find systematic correction factors that can be applied to reduce the bias between predicted and calculated absorbed doses. This concept was researched by Peters et al. [30] using pre-treatment 68Ga-PSMA-11 PET imaging to predict the absorbed dose in [177Lu]Lu-PSMA-617 treatment. They reported a scaling factor of 2.2 for kidneys. Lesions showed large interpatient variation. We could not determine such tentative PET/SPECT correction factors in this pilot study.

The predictive power of pre-treatment PET imaging has been studied more extensively in the case of somatostatin receptor radionuclide therapy with statistically significant yet modest results [31]. The kinetic difference between the short-lived PET tracers and the therapy agents ([18F]F-PSMA-1007 vs. [177Lu]Lu-PSMA-I&T [32]) may present a limitation that cannot be fully resolved. The short physical half-life of 18F limits the time-point of PET/CT imaging, giving an image of a relatively early distribution and clearance. A potential solution to address this issue could be using a longer-lived PET or SPECT-compatible isotope to enable imaging at later time-points [33, 34].

Conclusion

In this feasibility study, dose prediction for the first [177Lu]Lu-PSMA-I&T treatment cycle based on [18F]F-PSMA-1007 PET/CT overestimated kidney absorbed dose and underestimated bone lesion absorbed dose compared to post-treatment SPECT/CT-based absorbed doses. Observed substantial interpatient variation suggests caution in PET-based dose prediction and the need for further studies.

Acknowledgments

The authors declare no funding was received for this work.

NACP 2026 was financially supported by the Acta Oncologica Foundation.

Data availability statement

Research data are stored in an institutional repository and will be shared upon reasonable request to the corresponding author.

Ethics declarations & trial registry information

This study was conducted in accordance with the Declaration of Helsinki and its amendments. The study protocol was approved by the Institutional Review Board of Helsinki University Hospital (HUS/486/2025 and HUS/111/2026). As this was a retrospective study, the Institutional Review Board waived the requirement for patient informed consent, and no additional ethics statement was necessary.

Author contributions

This study bases on an extension of SK’s Master’s thesis, which was supervised by VR and EH, and reviewed by SS. VR designed and managed the study, including the administration of the research permission and acquiring patient data with the support of VA and MT. SPECT/CT phantom data were prepared by VR. SK performed all the quantitative analyses, result handling, and drafted the manuscript. JS and OS participated in finalizing the draft. All authors reviewed, edited, and approved the final manuscript.

References

[1]     Kratochwil C, Fendler WP, Eiber M, Hofman MS, Emmett L, Calais J, et al. Joint EANM/SNMMI procedure guideline for the use of 177Lu-labeled PSMA-targeted radioligand-therapy (177Lu-PSMA-RLT). Eur J Nucl Med Mol Imaging. 2023;50(9):2830–45. https://doi.org/10.1007/s00259-023-06255-8

[2]     Working group set up by the Finnish Medical Society Duodecim and the Finnish Urology Association. Prostate Cancer. Current care guidelines. Helsinki: The Finnish Medical Society Duodecim; 2025 [cited 2026 Jan 28]. Available from: www.kaypahoito.fi

[3]     Castellucci P, Deandreis D, Mirzaei S, Prior J, Sattler B, Schütze C. European Nuclear Medicine Guide: a joint publication by EANM and UEMS/EBNM. European Association of Nuclear Medicine EANM, 2026 edition. Vienna. 2026.

[4]     Stabin M. The practice of internal dosimetry in nuclear medicine. 1st ed. Boca Raton, FL: CRC Press; 2016.

[5]     Danieli R, Milano A, Gallo S, Veronese I, Lascialfari A, Indovinaet L, et al. Personalized dosimetry in targeted radiation therapy: a look to methods, tools and critical aspects. J Pers Med. 2022;12(2):205. https://doi.org/10.3390/jpm12020205

[6]     Sjögreen Gleisner K, Chouin N, Gabina PM, Cicone F, Gnesin S, Stokke C, et al. EANM dosimetry committee recommendations for dosimetry of 177Lu-labelled somatostatin-receptor- and PSMA-targeting ligands. Eur J Nucl Med Mol Imaging. 2022;49(6):1778–809. https://doi.org/10.1007/s00259-022-05727-7

[7]     Ells Z, Grogan TR, Czernin J, Dahlbom M, Calais J. Dosimetry of [177Lu]Lu-PSMA-targeted radiopharmaceutical therapies in patients with prostate cancer: a comparative systematic review and metaanalysis. J Nucl Med. 2024;65(8):1264–71. https://doi.org/10.2967/jnumed.124.267452

[8]     Tran-Gia J, Denis-Bacelar AM, Ferreira KM, Robinson AP, Calvert N, Fenwick AJ, et al. A multicentre and multi-national evaluation of the accuracy of quantitative Lu-177 SPECT/CT imaging performed within the MRTDosimetry project. EJNMMI Phys. 2021;8(1):55. https://doi.org/10.1186/s40658-021-00397-0

[9]     Farolfi A, Armstrong WR, Djaileb L, Gafita A, Hotta M, Allen-Auerbach M, et al. Differences and common ground in 177Lu-PSMA radioligand therapy practice patterns: international survey of 95 theranostic centers. J Nucl Med. 2024;65(3):438–45. https://doi.org/10.2967/jnumed.123.266391

[10]   Dickson J, Claeys W, Denis-bacelar A, Hristova I, Lassmann M, Koole M, et al. An EANM/EARL accreditation standard for quantitative Lu-177 SPECT-CT imaging. J Nucl Med. 2025;66(Supplement 1): 251783–251783.

[11]   Dadgar H, Karimzadeh R, Sharifi S, Hong X, Al balooshi B, Arabi H, et al. Predicting post-treatment outcomes of Lu-177 PSMA therapy using deep learning on pre-treatment Ga-68 PSMA PET images. J Nucl Med. 2025;66(Supplement 1):252132–252132.

[12]   Yazdani E, Neizehbaz A, Karamzade-Ziarati N, Emami F, Vosoughi H, Asadi M. Transforming [177Lu]Lu-PSMA-617 treatment planning: Machine learning-based radiodosiomics and swin UNETR using pretherapy PSMA positron emission tomography/computed tomography (PET/CT). Med Phys. 2025;52(10):e70030. https://doi.org/10.1002/mp.70030

[13]   Yazdani E, Sadeghi M, Karamzade-Ziarati N, Jabari P, Amini P, Vosoughi H. Machine learning-based dose prediction in [177Lu]Lu-PSMA-617 Therapy by integrating biomarkers and radiomic features from [68Ga]Ga-PSMA-11 Positron emission tomography/computed tomography. Int J Radiat Oncol Biol Phys. 2025;123(3):891–908. https://doi.org/10.1016/j.ijrobp.2025.05.014

[14]   Hohberg M, Reifegerst M, Drzezga A, Wild M, Schmidt M. Prediction of response to 177Lu-PSMA therapy based on tumor-to-kidney ratio on pretherapeutic PSMA PET/CT and posttherapeutic tumor-dose evaluation in mCRPC. J Nucl Med. 2023;64(11):1758–64. https://doi.org/10.2967/jnumed.122.264953

[15]   Violet J, Jackson P, Ferdinandus J, Sandhu S, Akhurst T, Iravani A, et al. Dosimetry of 177Lu-PSMA-617 in metastatic castration-resistant prostate cancer: correlations between pretherapeutic imaging and whole-body tumor dosimetry with treatment outcomes. J Nucl Med. 2019;60(4):517–23. https://doi.org/10.2967/jnumed.118.219352

[16]   Aide N, Lasnon C, Veit-Haibach P, Sera T, Sattler B, Boellaard R. EANM/EARL harmonization strategies in PET quantification: from daily practice to multicentre oncological studies. Eur J Nucl Med Mol Imaging 2017;44(Suppl 1):17–31. https://doi.org/10.1007/s00259-017-3740-2

[17]   Hudson HM, Larkin RS. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imaging. 1994;13(4):601–9. https://doi.org/10.1109/42.363108

[18]   Wan H. Automated contouring using neural networks. White Paper TD-650. Cleveland, OH: MIM Software Inc.; 2024.

[19]   Nelson AD, Nelson AS. MIM for therapy response evaluation. White Paper MIMTD-486. Cleveland, OH: MIM Software Inc.; 2018.

[20]   International Atomic Energy Agency (IAEA). Quality assurance for PET and PET/CT systems. IAEA Human Health Series No. 1. Vienna: International Atomic Energy Agency; 2009.

[21]   Reijonen V, Kokkonen S, Tenhunen M. Recovery coefficients in Lu-177-SPECT-CT are reduced as a function of activity concentration. EANM Innovation. 2025;1(Suppl):44–5. https://doi.org/10.1016/S3051-2913(25)00004-7

[22]   Madsen MT, Menda Y, O’Dorisio TM, O’Dorisio MS. Technical note: single time point dose estimate for exponential clearance. Med Phys. 2018;45(5):2318–24. https://doi.org/10.1002/mp.12886

[23]   Schuchardt C, Zhang J, Kulkarni HR, Chen X, Müller D, Baum RP. Prostate-specific membrane antigen radioligand therapy using 177Lu-PSMA I&T and 177Lu-PSMA-617 in patients with metastatic castration-resistant prostate cancer: comparison of safety, biodistribution, and dosimetry. J Nucl Med. 2022;63(8):1199–207. https://doi.org/10.2967/jnumed.121.262713

[24]   Karimzadeh A, Schatz L, Sauer M, Apostolova I, Buchert R, Klutmann S, et al. Organ and tumor dosimetry including method simplification for [177Lu]Lu-PSMA-I&T for treatment of metastatic castration resistant prostate cancer. EJNMMI Phys. 2024;11(1):63. https://doi.org/10.1186/s40658-024-00668-6

[25]   Okamoto S, Thieme A, Allmann J, D’Alessandria C, Maurer T, Retz M, et al. Radiation dosimetry for 177Lu-PSMA I&T in metastatic castration-resistant prostate cancer: absorbed dose in normal organs and tumor lesions. J Nucl Med. 2017;58(3):445–50. https://doi.org/10.2967/jnumed.116.178483

[26]   Feuerecker B, Chantadisai M, Allmann A, Tauber R, Allmann J, Steinhelfer L, et al. Pretherapeutic comparative dosimetry of 177Lu-rhPSMA-7.3 and 177Lu-PSMA I&T in patients with metastatic castration-resistant prostate cancer. J Nucl Med. 2022;63(6):833–9. https://doi.org/10.2967/jnumed.121.262671

[27]   Barna S, Haug AR, Hartenbach M, Rasul S, Grubmüller B, Kramer G, et al. Dose calculations and dose-effect relationships in 177Lu-PSMA I&T radionuclide therapy for metastatic castration-resistant prostate cancer. Clin Nucl Med. 2020;45(9):661–7. https://doi.org/10.1097/RLU.0000000000003157

[28]   Baum RP, Kulkarni HR, Schuchardt C, Singh A, Wirtz M, Wiessalla S, et al. 177Lu-labeled prostate-specific membrane antigen radioligand therapy of metastatic castration-resistant prostate cancer: safety and efficacy. J Nucl Med. 2016;57(7):1006–13. https://doi.org/10.2967/jnumed.115.168443

[29]   Hardiansyah D, Patrianesha BB, Glatting G. Population pharmacokinetic modeling in radiopharmaceutical therapy: a review. Front Nucl Med. 2025;5:1695332. https://doi.org/10.3389/fnume.2025.1695332

[30]   Peters SMB, Hofferber R, Privé BM, de Bakker M, Gotthardt M, Janssen M, et al. [68Ga]Ga-PSMA-11 PET imaging as a predictor for absorbed doses in organs at risk and small lesions in [177Lu]Lu-PSMA-617 treatment. Eur J Nucl Med Mol Imaging. 2022;49(4):1101–12. https://doi.org/10.1007/s00259-021-05538-2

[31]   Akhavanallaf A, Lu Z, Peterson AB, Blakkisrud J, Kurkowska S, Yadav S, et al. Can 177Lu-DOTATATE kidney absorbed doses be predicted from pretherapy SSTR PET? Findings from multicenter data. J Nucl Med. 2025;66(7):1082–90. https://doi.org/10.2967/jnumed.124.269098

[32]   Capasso G, Stefanucci A, Tolomeo A. A systematic review on the current status of PSMA-targeted imaging and radioligand therapy. Eur J Med Chem. 2024;263:115966. https://doi.org/10.1016/j.ejmech.2023.115966

[33]   Kabasakal L, AbuQbeitah M, Aygün A,Yeyin N, Ocak M, Demirci E, et al. Pre-therapeutic dosimetry of normal organs and tissues of (177)Lu-PSMA-617 prostate-specific membrane antigen (PSMA) inhibitor in patients with castration-resistant prostate cancer. Eur J Nucl Med Mol Imaging. 2015;42(13):1976–83. https://doi.org/10.1007/s00259-015-3125-3

[34]   Privé BM, Derks YHW, Rosar F, Franssen GM, Peters SMB, Khreish F, et al. 89Zr-labeled PSMA ligands for pharmacokinetic PET imaging and dosimetry of PSMA-617 and PSMA-I&T: a preclinical evaluation and first in man. Eur J Nucl Med Mol Imaging. 2022;49(6):2064–76. https://doi.org/10.1007/s00259-021-05661-0