Dosimetric quantification of the incidental irradiation of the ‘true’ (deep) ano-inguinal lymphatic drainage of anal cancer patients not described in conventional contouring guidelines

Authors

  • Hendrik Dapper Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Gregor Habl Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Christoph Hirche Department for Hand-, Plastic and Rekonstructive Surgery, Burn Centre BG-Trauma Centre Ludwigshafen/RhineBG – Klinik Ludwigshafen, University of Heidelberg, Ludwigshafen, Germany
  • Stefan Münch Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Markus Oechsner Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Michael Mayinger Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Christina Sauter Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany
  • Stephanie E. Combs Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany; ;Institute for Innovative Radiotherapie (iRT), Helmholtz Zentrum München, Neuherberg, Germany; ;Deutsches Konsortium für Translationale Krebsforschung (DKTK), Munich, Germany
  • Daniel Habermehl Department of Radiation Oncology, Klinikum Rechts der Isar, TU München, Munich, Germany; ;Institute for Innovative Radiotherapie (iRT), Helmholtz Zentrum München, Neuherberg, Germany

DOI:

https://doi.org/10.1080/0284186X.2017.1415459

Abstract

Introduction: The ano-inguinal lymphatic drainage (AILD) is located in the subcutaneous adipose tissue of the proximal medial thigh. Findings from fluorescence methods give us new information about anatomical conditions of the AILD. Current contouring guidelines do not advise the inclusion of the ‘true’ AILD into the clinical target volume (CTV). Aim of this work was the retrospective analysis of the incidental dose to the AILD in an anal cancer (AC) patient cohort who underwent definitive chemoradiation (CRT) therapy with Volumetric Arc Therapy – Intensity Modulated Radiation Therapy (VMAT-IMRT).

Methods: VMAT-IMRT plans of 15 AC patients were analyzed. Based on findings from new fluorescence methods we created a new volume, the expected AILD. The examined dosimetric parameters were the minimal, maximal and mean dose and V10–V50 that were delivered to the AILD, respectively.

Results: The median volume of AILD was 1047 cm³. Mean DminDmax and Dmean were 7.5 Gy, 58.9 Gy and 40.8 Gy for AILD. The clinical relevant dose of 30.0 Gray covered in mean 76% of the volume of the AILD, respectively.

Conclusions: Only 76% of the AILD-volume received at least an expected required treatment dose of 30 Gy incidentally. Concerning the low number of loco-regional relapses in AC patients after definitive CRT one has to balance increased side effects against a rigid oncological–anatomical interpretation of the local lymphatic drainage by including the AILD into the standard CTV.

Downloads

Download data is not yet available.

Downloads

Published

2018-06-03

How to Cite

Dapper, H. ., Habl, G. ., Hirche, C. ., Münch, S. ., Oechsner, M. ., Mayinger, M. ., … Habermehl, D. . (2018). Dosimetric quantification of the incidental irradiation of the ‘true’ (deep) ano-inguinal lymphatic drainage of anal cancer patients not described in conventional contouring guidelines. Acta Oncologica, 57(6), 825–830. https://doi.org/10.1080/0284186X.2017.1415459