Real time transit dosimetry for the breath-hold radiotherapy technique: An initial experience

Authors

  • Angelo Piermattei Istituto di Fisica, Università Cattolica S. Cuore, Roma, Italy; U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Savino Cilla U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Luca Grimaldi U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Pietro Viola U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Lorenzo Frattarolo U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Guido D'Onofrio U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Maurizio Craus U.O. Fisica Sanitaria, Università Cattolica S. Cuore, Campobasso, Italy
  • Andrea Fidanzio Istituto di Fisica, Università Cattolica S. Cuore, Roma, Italy
  • Luigi Azario Istituto di Fisica, Università Cattolica S. Cuore, Roma, Italy
  • Francesca Greco Istituto di Fisica, Università Cattolica S. Cuore, Roma, Italy
  • Cinzia Digesu U.O. Radioterapia, Università Cattolica S. Cuore, Campobasso, Italy
  • Francesco Deodato U.O. Radioterapia, Università Cattolica S. Cuore, Campobasso, Italy
  • Gabriella Macchia U.O. Radioterapia, Università Cattolica S. Cuore, Campobasso, Italy
  • Alessio G. Morganti U.O. Radioterapia, Università Cattolica S. Cuore, Campobasso, Italy

DOI:

https://doi.org/10.1080/02841860802304572

Abstract

Introduction. The breath-hold is one of the techniques to obtain the dose escalation for lung tumors. However, the change of the patient's breath pattern can influence the stability of the inhaled air volume, IAV, used in this work as a surrogate parameter to assure the tumor position reproducibility during dose delivery.Materials. and methodIn this paper, an Elekta active breathing coordinator has been used for lung tumor irradiation. This device is not an absolute spirometer and the feasibility study here presented developed (i) the possibility to select a specific range ε of IAV values comfortable for the patient and (ii) the ability of a transit signal rate , obtained by a small ion-chamber positioned on the portal image device, to supply in real time the in vivo isocenter dose reproducibility. Indeed, while the selection of the IAV range depends on the patient's ability to follow instructions for breath-hold, the  monitoring can supply to the radiation therapist a surrogate of the tumor irradiation reproducibility.Results. The detection of the  in real time during breath-hold was used to determine the interfraction isocenter dose variations due to the reproducibility of the patient's breathing pattern. The agreement between the reconstructed and planned isocenter dose in breath-hold at the interfraction level was well within 1.5%, while in free breathing a disagreement up to 8% was observed. The standard deviation of the  in breath-hold observed at the intrafraction level is a bit higher than the one obtained without the patient and this can be justified by the presence of a small residual tumor motion as heartbeat.Conclusion. The technique is simple and can be implemented for routine use in a busy clinic.