Two-dimensional in vivo rectal dosimetry during high-dose-rate brachytherapy for cervical cancer: a phantom study

Authors

  • Eun Hee Jeang Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Youngmoon Goh Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Kwan Ho Cho Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Soonki Min Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Sang Hyoun Choi Department of Radiation Oncology, Korea Cancer Center Hospital, Seoul, Korea
  • Hojin Jeong Department of Radiation Oncology, Gyeongsang National University Hospital, Jinju, Korea
  • Kwanghyun Jo Department of Radiation Oncology, Samsung Medical Center, Seoul, Korea
  • Nuri Lee Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Sanghyeon Song Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Se Byeong Lee Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Dongho Shin Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Joo-Young Kim Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Dae Yong Kim Proton Therapy Center, National Cancer Center, Goyang, Korea
  • Ui-Jung Hwang Department of Radiation Oncology, National Medical Center, Seoul, Korea
  • Young Kyung Lim Proton Therapy Center, National Cancer Center, Goyang, Korea

DOI:

https://doi.org/10.1080/0284186X.2018.1484155

Abstract

Background: The aim of the present study was to verify the dosimetric accuracy of two-dimensional (2D) in vivo rectal dosimetry using an endorectal balloon (ERB) with unfoldable EBT3 films for high-dose-rate (HDR) brachytherapy for cervical cancer. The clinical applicability of the technique was discussed.

Material and methods: ERB inflation makes the EBT3 films unrolled, whereas its deflation makes them rolled. Patient-specific quality assurance (pQA) tests were performed in 20 patient plans using an Ir-192 remote afterloading system and a water-filled cervical phantom with the ERB. The dose distributions measured in ERBs were compared with those of the treatment plans.

Results: The absolute dose profiles measured by the ERBs were in good agreement with those of treatment plans. The global gamma passing rates were 96–100% and 91–100% over 20 pQAs under the criteria of 3%/3 mm and 3%/2 mm, respectively, with a 30% low-dose threshold. Dose-volume histograms of the rectal wall were obtained from the measured dose distributions and showed small volume differences less than 2% on average from the patients’ plans over the entire dose interval. The positioning error of the applicator set was detectable with high sensitivity of 12% dose area variation per mm. Additionally, the clinical applicability of the ERB was evaluated in volunteers, and none of them felt any pain when the ERB was inserted or removed.

Conclusions: The 2D in vivo rectal dosimetry using the ERB with EBT3 films was effective and might be clinically applicable for HDR brachytherapy for cervical and prostate cancers to monitor treatment accuracy and consistency as well as to predict rectal toxicity.

 

Downloads

Download data is not yet available.

Downloads

Published

2018-10-03

How to Cite

Hee Jeang, E. ., Goh, Y. ., Ho Cho, K. ., Min, S., Hyoun Choi, S. ., Jeong, H. ., … Kyung Lim, Y. (2018). Two-dimensional in vivo rectal dosimetry during high-dose-rate brachytherapy for cervical cancer: a phantom study. Acta Oncologica, 57(10), 1359–1366. https://doi.org/10.1080/0284186X.2018.1484155