Breast irradiation with three conformal photon fields for patients with high lung involvement

Authors

  • L. Cozzi From the Oncology Institute of Southern Switzerland, Medical Physics Unit
  • A. Fogliata From the Oncology Institute of Southern Switzerland, Medical Physics Unit
  • G. Nicolini From the Oncology Institute of Southern Switzerland, Medical Physics Unit
  • T. Rancati Radiation Oncology (J. Bernier), Bellinzona, Switzerland, and the Department of Physics, University of Milan,Italy

DOI:

https://doi.org/10.1080/02841860410015280

Abstract

Since 2001, 50 breast cancer patients, for whom extensive lung/heart involvement was expected from the use of conventional tangential 2-field technique (2F) owing to complex anatomies, were irradiated using a 3-field conformal technique (3F). Dose plans were designed for both 3F and 2F and a dose volume histogram analysis on ipsilateral lung, heart, and target was conducted. The 3F technique allowed a significant reduction in ipsilateral lung irradiation: mean dose dropped from 16.0±3.8 (2F) to 12.0±2.7 Gy (3F) and V45Gy from 20.7±6.8 (2F) to 3.2±2.9% (3F). Similarly, in patients irradiated to the left breast, heart mean dose was reduced from 8.1 Gy (2F) to 6.8 Gy (3F) and D15% from 19.0 Gy to 14.0 Gy. All differences reached a high degree of significance. The target coverage was not clinically compromised since the slight reduction using 3F compared with 2F is limited to V95% while V90% difference, even if statistically significant, is small: 98.2±1.8% (3F) and 98.8±1.6 (2F). A preliminary report on clinical follow up is also included; with a mean follow up of 15.8 months, no pulmonary or cardiac complications were observed.

 

Downloads

Download data is not yet available.

Downloads

Published

2004-09-01

How to Cite

Cozzi, L. ., Fogliata, A. ., Nicolini, G. ., & Rancati, T. . (2004). Breast irradiation with three conformal photon fields for patients with high lung involvement. Acta Oncologica, 43(6), 558–566. https://doi.org/10.1080/02841860410015280