Wedge Filters for Megavoltage Roentgen Ray Beams

Authors

  • R Van der Laarse Radiation Therapy Department, Antoni Van Leeuwenhoek Hospital, the Netherlands Cancer Institute, Amsterdam, the Netherlands; Philips Medical Systems Division, Best, and the Laboratory of Medical Physics, University of Amsterdam, Amsterdam, the Netherlands
  • P. A. M van Overbeek Radiation Therapy Department, Antoni Van Leeuwenhoek Hospital, the Netherlands Cancer Institute, Amsterdam, the Netherlands; Philips Medical Systems Division, Best, and the Laboratory of Medical Physics, University of Amsterdam, Amsterdam, the Netherlands
  • J Strackee Radiation Therapy Department, Antoni Van Leeuwenhoek Hospital, the Netherlands Cancer Institute, Amsterdam, the Netherlands; Philips Medical Systems Division, Best, and the Laboratory of Medical Physics, University of Amsterdam, Amsterdam, the Netherlands

DOI:

https://doi.org/10.3109/02841868409136052

Abstract

The aim of this investigation, to construct a range of fixed wedge filters and to simulate these with a motorized wedge, led to the derivation of 5 equations. These equations can be used to construct and test a consistent set of fixed wedge filters, eliminating elaborate trial and error experiments. The fixed wedge filters already in existence for the Philips SL75–10 and SL75–20 linear accelerators fitted these equations rather well. After adapting the motorized wedge of the SL75–14 according to these equations, it simulated the fixed wedges accurately.

Downloads

Download data is not yet available.

Downloads

Published

1984-01-01

How to Cite

Van der Laarse, R., van Overbeek, P. A. M., & Strackee, J. (1984). Wedge Filters for Megavoltage Roentgen Ray Beams. Acta Oncologica, 23(6), 477–484. https://doi.org/10.3109/02841868409136052