Properties of paraffin wax as a bolus material in accelerator-based boron neutron capture therapy
DOI:
https://doi.org/10.2340/1651-226X.2025.44024Keywords:
BNCT, AB-BNCT, Bolus, Dosimetry, Paraffin waxAbstract
Background and purpose: Boron neutron capture therapy (BNCT) is targeted radiation therapy enabling cellular-level cancer treatment. With epithermal neutrons, the dose maximum typically occurs ~2 cm deep in tissue, challenging superficial tumor control. As in external beam radiation therapy, surface dose can be increased using a bolus. However, in BNCT, tissue equivalency is complex and strongly dependent on elemental composition. This study examined a paraffin wax bolus’s effect on the epithermal neutron beam in accelerator-based BNCT (AB-BNCT) and evaluated agreement between treatment planning system (TPS) calculations and measurements.
Materials and methods: Beam characterization used the neutron activation method with gold and manganese foils. Due to its high cross-section for thermal neutrons, manganese activation serves as a surrogate for boron dose estimation. Irradiations were conducted in a 3D water tank and in a head-shaped phantom with 5 and 10 mm boluses. Dose calculation utilized the newly commissioned RayStation TPS with a Monte Carlo-based engine built on the GEANT4 toolkit.
Results: Calculated and measured results agree within 5% accuracy in significant dose region (>50% dose). Near the surface and at greater depths, agreement remains within 10%. The bolus shifts the activation depth curve toward the surface by 4–13 mm depending on its thickness. Manganese surface activation increases from 30% without a bolus to ~70% and ~ 90% with 5 and 10 mm boluses, respectively.
Interpretation: Paraffin wax effectively moderates neutron energy, making it a suitable bolus material for AB-BNCT treatments requiring increased surface dose.
Downloads
References
Savolainen S, Kortesniemi M, Timonen M, Reijonen V, Kuusela L, Uusi-Simola J, et al. Boron neutron capture therapy (BNCT) in Finland: technological and physical prospects after 20 years of experiences. Phys Medica. 2013;29(3):233–48.
https://doi.org/10.1016/j.ejmp.2012.04.008 DOI: https://doi.org/10.1016/j.ejmp.2012.04.008
Jin WH, Seldon C, Butkus M, Sauerwein W, Giap HB. A review of boron neutron capture therapy: its history and current challenges. Int J Part Ther. 2022;9(1):71–82.
https://doi.org/10.14338/IJPT-22-00002.1 DOI: https://doi.org/10.14338/IJPT-22-00002.1
Goldstone KE. Tissue substitutes in radiation dosimetry and measurement, in: ICRU Report 44. International Commission on Radiation Units and Measurements, Bethesda, MD (United States), 1989.
Porra L, Wendland L, Seppälä T, Koivunoro H, Revitzer H, Tervonen J, et al. From nuclear reactor-based to proton accelerator-based therapy: the Finnish boron neutron capture therapy experience. Cancer Biother Radiopharm. 2023;38(3):184–91.
https://doi.org/10.1089/cbr.2022.0059 DOI: https://doi.org/10.1089/cbr.2022.0059
Seppälä T, Collan J, Auterinen I, Serén T, Salli E, Kotiluoto, et al. A dosimetric study on the use of bolus materials for treatment of superficial tumors with BNCT. Appl Radiat Isot. 2004;61(5):787–91.
https://doi.org/10.1016/j.apradiso.2004.05.054 DOI: https://doi.org/10.1016/j.apradiso.2004.05.054
Hubbell J, Seltzer S. Tables of X-Ray Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients (version 1.4), (2004). National Institute of Standards and Technology, Gaithersburg, MD, (United States), http://physics.nist.gov/xaamdi.
Sasaki A, Tanaka H, Takata T, Tamari Y, Watanabe T, Hu N, et al. Development of an irradiation method for superficial tumours using a hydrogel bolus in an accelerator-based BNCT. Biomed Phys Eng Express. 2021;8(1):015015.
https://doi.org/10.1088/2057-1976/ac3d73 DOI: https://doi.org/10.1088/2057-1976/ac3d73
IAEA. Handbook on nuclear activation cross-sections. Technical Reports Series No. 156. Vienna: IAEA; 1974.
Wendland L, Porra L, Seppälä T, Tenhunen M. Neutron activation in interrupted neutron beams. Appl Radiat Isot 2025;225:118896.
https://doi.org/10.1016/j.apradiso.2025.111896 DOI: https://doi.org/10.1016/j.apradiso.2025.111896
Porra L, Seppälä T, Wendland L, Revitzer H, Joensuu H, Eide P, et al. Accelerator-based boron neutron capture therapy facility at the Helsinki University Hospital. Acta Oncol. 2022;61(2):269–73.
https://doi.org/10.1080/0284186X.2021.1979646 DOI: https://doi.org/10.1080/0284186X.2021.1979646
GEANT4 A Simulation Toolkit [Internet]. version 11.1. [Cited 1.3.2025] Available from: https://geant4.web.cern.ch/.
ROOT Data Analysis Framework [Internet]. version 6.34.02. [Cited 1.2.2025] Available from: https://root.cern.ch/.
James F. MINUIT: function minimization and error analysis reference manual. Technical report. CERN Geneva, Switzerland; 1998.
Compositions of materials used in STAR databases. National Institute of Standards and Technology, Gaithersburg, MD, (United States), https://physics.nist.gov/cgi-bin/Star/compos.pl.
Malik A, Ogden S, Amberg G, Hjort K. Modeling and analysis of a phase change material thermohydraulic membrane microactuator. J Microelectromech Syst. 2012;22(1):186–94.
https://doi.org/10.1109/JMEMS.2012.2222866 DOI: https://doi.org/10.1109/JMEMS.2012.2222866
Published
How to Cite
Issue
Section
Categories
License
Copyright (c) 2025 Jenna Tarvonen; Lauri Wendland; Liisa Porra, Tiina Seppälä, Mikko Tenhunen

This work is licensed under a Creative Commons Attribution 4.0 International License.
