Reproducibility of goniometry of the wrist
DOI:
https://doi.org/10.2340/1650197719861857Abstract
The wrist motion in 31 healthy probands was measured by four observers with a simple goniometer under standardized conditions. The INTRA-observer variation was calculated to 5 to 8 degrees and the INTER-observer variation to 6 to 10 degrees. No difference between experienced and non-experienced observers was found. The difference between the right and the left wrist was negligible, indicating that the opposite wrist can be used as reference when evaluating restrictions in motion.Downloads
Downloads
Published
How to Cite
Issue
Section
License
All digitalized JRM contents is available freely online. The Foundation for Rehabilitation Medicine owns the copyright for all material published until volume 40 (2008), as from volume 41 (2009) authors retain copyright to their work and as from volume 49 (2017) the journal has been published Open Access, under CC-BY-NC licences (unless otherwise specified). The CC-BY-NC licenses allow third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material for non-commercial purposes, provided proper attribution to the original work.