A computerized wheelchair ergometer. Results of a comparison study

Authors

  • van der Woude LH Veeger HE

DOI:

https://doi.org/10.2340/16501977921723

Abstract

To determine the validity of propulsion simulation on a stationary wheelchair ergometer, nine male able-bodied subjects performed submaximal exercise tests on the ergometer and on a motor driven treadmill (MDT). Oxygen uptake, ventilation and stroke parameters were equal for both devices, but heart rate was lower and trunk movement was less for the ergometer test. Analysis of forces and power output on the ergometer indicated that power output was equal for both wheels. The ratio between applied forces and the effectively directed force component was approximately 80%. Also a small torque was applied by the hand onto the handrim surface which contributed to the total propulsion torque around the axle. It is concluded that the ergometer is capable of simulation of wheelchair propulsion, although the different trunk motion may necessitate sufficient wheelchair propulsion experience. Force analysis results are discussed.

Downloads

Download data is not yet available.

Downloads

Published

1992-01-01

How to Cite

Veeger HE, van der W. L. (1992). A computerized wheelchair ergometer. Results of a comparison study. Journal of Rehabilitation Medicine, 24(1), 17–23. https://doi.org/10.2340/16501977921723