Validation of heart rate measurements from the polar verity sense during the submaximal Ekblom-Bak test in adult individuals with obesity

Authors

  • Grete Rudi Onstad Department of Neuromedicine and Movement Science, Faculty of Medicine and Health Sciences, NTNU Norwegian University of Science and Technology, Trondheim, Norway
  • Fredrik Klæboe Lohne Department of Neuromedicine and Movement Science, Faculty of Medicine and Health Sciences, NTNU Norwegian University of Science and Technology, Trondheim, Norway; National Research Centre for the Working Environment, Copenhagen, Denmark
  • Marius Steiro Fimland Department of Neuromedicine and Movement Science, Faculty of Medicine and Health Sciences, NTNU Norwegian University of Science and Technology, Trondheim, Norway; Unicare Helsefort Rehabilitation Centre, Rissa, Norway

DOI:

https://doi.org/10.2340/jrm.v58.44975

Keywords:

Obesity, heart rate, photoplethysmogram, Polar Verity Sense, Polar H10, Ekblom-Bak cycle test, Validation, Feasibility

Abstract

Objective: To assess the validity of Polar Verity Sense, an arm-worn heart rate (HR) monitor, for measuring HR in individuals with obesity during the submaximal ergometer cycle Ekblom-Bak test.

Subjects/Patients: Adults in an inpatient rehabilitation programme for individuals with obesity.

Methods: Participants completed the Ekblom-Bak test wearing the Polar H10 chest strap and 2 Verity Sense devices on the forearm and upper arm. Statistical analyses included intraclass correlation coefficient (ICC), Lin’s concordance correlation coefficient (CCC), Bland–Altman analyses with 95% limits of agreement (LoA), and mean absolute percentage error (MAPE). Feasibility was assessed with a post-test questionnaire.

Results: 49 participants were included in the analyses of estimated maximal oxygen consumption (VO2max), while 48 were included in analyses of HR data. ICC and CCC values ranged from 0.998–0.999 between the H10 chest strap and Verity Sense measures. Small differences were observed in Bland–Altman plots. MAPE ranged from 0.46% to 0.51%. Feasibility responses from the questionnaire showed that participants preferred the Verity Sense to the H10.

Conclusion: Polar Verity Sense accurately measures HR and estimates VO2max in adults with obesity during the Ekblom-Bak test. Individuals with obesity may freely opt for either device without sacrificing accuracy during steady-state activities at low and moderate intensities. This choice may eliminate a barrier in assessing their cardiorespiratory fitness.

Downloads

Download data is not yet available.

References

Mayoral LP, Andrade GM, Mayoral EP, Huerta TH, Canseco SP, Rodal Canales FJ, et al. Obesity subtypes, related biomarkers & heterogeneity. Indian J Med Res 2020; 151: 11–21. DOI: https://doi.org/10.4103/ijmr.IJMR_1768_17

Kaminsky LA, Arena R, Ellingsen Ø, Harber MP, Myers J, Ozemek C, et al. Cardiorespiratory fitness and cardiovascular disease: the past, present, and future. Prog Cardiovasc Dis 2019; 62: 86–93. DOI: https://doi.org/10.1016/j.pcad.2019.01.002

Ross R, Blair SN, Arena R, Church TS, Després JP, Franklin BA, et al. Importance of assessing cardiorespiratory fitness in clinical practice: a case for fitness as a clinical vital sign: a scientific statement from the American Heart Association. Circulation 2016; 134: e653–e699. DOI: https://doi.org/10.1161/CIR.0000000000000461

Björkman F, Ekblom-Bak E, Ekblom Ö, Ekblom B. Validity of the revised Ekblom Bak cycle ergometer test in adults. Eur J Appl Physiol 2016; 116: 1627–1638. DOI: https://doi.org/10.1007/s00421-016-3412-0

Arena R, Myers J, Williams MA, Gulati M, Kligfield P, Balady GJ, et al. Assessment of functional capacity in clinical and research settings: a scientific statement from the American Heart Association Committee on Exercise, Rehabilitation, and Prevention of the Council on Clinical Cardiology and the Council on Cardiovascular Nursing. Circulation 2007; 116: 329–343. DOI: https://doi.org/10.1161/CIRCULATIONAHA.106.184461

Gilgen-Ammann R, Schweizer T, Wyss T. RR interval signal quality of a heart rate monitor and an ECG Holter at rest and during exercise. Eur J Appl Physiol 2019; 119: 1525–1532. DOI: https://doi.org/10.1007/s00421-019-04142-5

Schubert MM, Clark A, De La Rosa AB. The Polar (®) OH1 optical heart rate sensor is valid during moderate-vigorous exercise. Sports Med Int Open 2018; 2: E67–e70. DOI: https://doi.org/10.1055/a-0631-0920

Hettiarachchi IT, Hanoun S, Nahavandi D, Nahavandi S. Validation of Polar OH1 optical heart rate sensor for moderate and high intensity physical activities. PLoS One 2019; 14: e0217288. DOI: https://doi.org/10.1371/journal.pone.0217288

Temko A. Accurate heart rate monitoring during physical exercises using PPG. IEEE Trans Biomed Eng 2017; 64: 2016–2024. DOI: https://doi.org/10.1109/TBME.2017.2676243

Singh N, Moneghetti KJ, Christle JW, Hadley D, Plews D, Froelicher V. Heart rate variability: an old metric with new meaning in the era of using mHealth technologies for health and exercise training guidance. Part One: Physiology and methods. Arrhythm Electrophysiol Rev 2018; 7: 193–198. DOI: https://doi.org/10.15420/aer.2018.27.2

Muggeridge DJ, Hickson K, Davies AV, Giggins OM, Megson IL, Gorely T, et al. Measurement of heart rate using the Polar OH1 and Fitbit Charge 3 wearable devices in healthy adults during light, moderate, vigorous, and sprint-based exercise: validation study. JMIR Mhealth Uhealth 2021; 9: e25313. DOI: https://doi.org/10.2196/25313

Navalta JW, Davis DW, Malek EM, Carrier B, Bodell NG, Manning JW, et al. Heart rate processing algorithms and exercise duration on reliability and validity decisions in biceps-worn Polar Verity Sense and OH1 wearables. Sci Rep 2023; 13: 11736. DOI: https://doi.org/10.1038/s41598-023-38329-w

López-Belmonte Ó, Febles-Castro A, Gay A, Cuenca-Fernández F, Arellano R, Ruiz-Navarro JJ. Validity of the polar verity sense during swimming at different locations and intensities. Scand J Med Sci Sports 2023. DOI: https://doi.org/10.1111/sms.14494

Schweizer T, Gilgen-Ammann R. Wrist-worn and arm-worn wearables for monitoring heart rate during sedentary and light-to-vigorous physical activities: device validation study. JMIR Cardio 2025; 9: e67110. DOI: https://doi.org/10.2196/67110

Schneider C, Hanakam F, Wiewelhove T, Döweling A, Kellmann M, Meyer T, et al. Heart rate monitoring in team sports: a conceptual framework for contextualizing heart rate measures for training and recovery prescription. Front Physiol 2018; 9: 639. DOI: https://doi.org/10.3389/fphys.2018.00639

Unicare. Sykelig overvekt voksen: Unicare Helsefort; 2024. Available from: https: //unicare.no/helsefort/rehabiliteringstilbud/sykelig-overvekt-voksen/

Swedish School of Sport and Health Science. The Ekblom-Bak test: GIH; 2024 [updated 12. February 2024. Available from: https: //www.gih.se/english/research/laboratories/the-astrand-laboratory/the-ekblom-bak-test

Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud 2010; 47: 931–936. DOI: https://doi.org/10.1016/j.ijnurstu.2009.10.001

Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007; 17: 571–582. DOI: https://doi.org/10.1080/10543400701329422

Reddy RK, Pooni R, Zaharieva DP, Senf B, El Youssef J, Dassau E, et al. Accuracy of wrist-worn activity monitors during common daily physical activities and types of structured exercise: evaluation study. JMIR Mhealth Uhealth 2018; 6: e10338. DOI: https://doi.org/10.2196/10338

Shcherbina A, Mattsson CM, Waggott D, Salisbury H, Christle JW, Hastie T, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med 2017; 7. DOI: https://doi.org/10.3390/jpm7020003

Navalta JW, Montes J, Bodell NG, Salatto RW, Manning JW, DeBeliso M. Concurrent heart rate validity of wearable technology devices during trail running. PLoS One 2020; 15: e0238569. DOI: https://doi.org/10.1371/journal.pone.0238569

Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016; 15: 155–163. DOI: https://doi.org/10.1016/j.jcm.2016.02.012

Olstad BH, Zinner C. Validation of the Polar OH1 and M600 optical heart rate sensors during front crawl swim training. PLoS One 2020; 15: e0231522. DOI: https://doi.org/10.1371/journal.pone.0231522

Mühlen JM, Stang J, Lykke Skovgaard E, Judice PB, Molina-Garcia P, Johnston W, et al. Recommendations for determining the validity of consumer wearable heart rate devices: expert statement and checklist of the INTERLIVE Network. Br J Sports Med 2021; 55: 767–779. DOI: https://doi.org/10.1136/bjsports-2020-103148

Downloads

Additional Files

Published

2026-09-07

How to Cite

Onstad, G. R., Lohne , F. K., & Fimland , M. S. (2026). Validation of heart rate measurements from the polar verity sense during the submaximal Ekblom-Bak test in adult individuals with obesity. Journal of Rehabilitation Medicine, 58, jrm44975. https://doi.org/10.2340/jrm.v58.44975

Issue

Section

Original Report

Categories