TEWL measurement standardization: kinetic and topographic aspects.
DOI:
https://doi.org/10.2340/0001555574171175Abstract
The aim of this study was to standardize the recording technique of the SERVO MED Ep 1C to determine the rest time required for human subjects to obtain stabilized values of transepidermal water loss (TEWL). Measurements of TEWL were performed on the flexor side of the forearm in healthy volunteers. First, different sites were tested on different days to study intra- and inter-individual variations. The recordings were carried out every 5 min for 2 h. Values reached a linear state after 15 min of rest. A time-course type curve of TEWL was established.Downloads
Downloads
Published
How to Cite
Issue
Section
License
All digitalized ActaDV contents is available freely online. The Society for Publication of Acta Dermato-Venereologica owns the copyright for all material published until volume 88 (2008) and as from volume 89 (2009) the journal has been published fully Open Access, meaning the authors retain copyright to their work.
Unless otherwise specified, all Open Access articles are published under CC-BY-NC licences, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material for non-commercial purposes, provided proper attribution to the original work.