Surface characterization analysis of failed dental implants using scanning electron microscopy

Authors

  • Umer Daood International Medical University, Malaysia
  • Ninette Bandey UAE Health Authority, United Arab Emirates
  • Saad Bin Qasim King Saud University, Riyadh
  • Hanan Omar International Medical University, Malaysia
  • Saad A. Khan International Medical University, Malaysia

DOI:

https://doi.org/10.3109/00016357.2011.569507

Keywords:

titanium implants, dental implants, surface morphology

Abstract

Objective . To investigate the failure of 15 dental implants (Paragon/Zimmer) in relation to their surface quality. Materials and methods . The study comprised of 15 dental implants (7 mm D Advent Implant, 3.9 mm D apex design implant), which were followed from surgery to completion of prosthetic restorations. The implants were placed during a 6-year period from 2003–2009 in non-smoking patients (male; 7, females; 5). There were eight upper and seven lower implants. Surface characterization after immersion in SBF of these failed implants was investigated using SEM and EDS compared to that of an unused implant of the same brand. Results. Results revealed that, following immersion in SBF, the implant surfaces showed new components like Ca+, Na+ and Cl, but in trace quantities. Conclusions . After SEM observation and EDS analysis, it was concluded that the apatite layer formation could not be verified.